Fast GC의 분석성능평가와 실내환경의 VOC 분석: 시료의 주입방식별 중량특성 비교
(주)코리아스칼라
- 최초 등록일
- 2016.04.01
- 최종 저작일
- 2008.06
- 17페이지/
어도비 PDF
- 가격 5,100원
![할인쿠폰받기](/images/v4/document/ico_det_coupon.gif)
* 본 문서는 배포용으로 복사 및 편집이 불가합니다.
서지정보
ㆍ발행기관 : 한국실내환경학회
ㆍ수록지정보 : 한국실내환경학회지 / 5권 / 2호
ㆍ저자명 : 안지원, 김기현
영어 초록
In this study, the performance of fast gas chromatography system was evaluated using VOC standards prepared in both liquid and gaseous phase. When the liquid‐phase VOC standards were analyzed by both direct injection and HS‐SPME method, all the chromatographic separation was completed within 4 minutes. The calibration experiments were conducted further using gaseous standard of BTX. The calibration results derived by direct injection method generally showed good linearities, regardless of phases, while it was not the case for HS‐SPME method. In case of liquid‐phase standard, MDL values for direct injection and HSSPME method were calculated in the range of 0.07~0.19 to 0.63~3.76 ng, respectively. In contrast, MDL values for gaseous standard were 0.27~0.45 and 1.94~6.90, respectively. The reproducibility of our method, when expressed in terms RSE, showed above 5 %. When the sensitivity of different techniques is compared, the calibration slope values of BTX decreased on the order: direct injection of liquid‐phase standard > HSSPME method of liquid‐phase standard > direct injection of gaseous standard > HS‐SPME method of gaseous standard. Although fast GC is very efficient to reduce the total running time significantly, extended studies are desirable to improve its reproducibility.